Particle size distributions at process or end-use conditions

Poster presented at Mini-APACT 2026, AstraZeneca, Macclesfield, UK, 24–25 September 2026.

Summary

Sizing from tens of nanometres to several microns in a single acquisition using the VISTA algorithm.

A process analytic is only useful if it reports the state of the process, not the state of the sample after it has been prepared for the instrument. Most particle sizing techniques need the sample brought into a narrow optical working window, which for a concentrated process stream usually means dilution. That is not a neutral operation: self-assembled structures can be lost, osmotic changes can drive growth, and weakly bound early aggregates can be broken.

This poster sets out measurements made with Lumero S1, which records a short microscopy video of the sample in a disposable chip, using as little as 3 µL. VISTA (Video Intensity Spatio-Temporal Analysis) analyses the image intensity fluctuations arising from Brownian motion of the particles over a range of length and time scales, and returns a distribution of diffusivities, converted to hydrodynamic radius through the Stokes–Einstein relation. Because the analysis is statistical rather than single-particle, particles need not be resolved or tracked, and billions can be captured in one measurement.

Single frame from a Lumero acquisition of a lentiviral vector sample, shown beside its intensity-weighted and volume-weighted size distributions which reveal coexisting small and large particles.
Figure 1 (Fig. 1 in the poster). A single frame from a Lumero acquisition. Only the largest particles are individually visible; the sub-resolution population still modulates the image intensity and is measured by VISTA even though it is invisible to manual inspection, and the spatial localisation of large particle signal means the small signal can still be detected.
Intensity-weighted particle size distribution of lipid nanoparticles measured by DLS and by Lumero, showing a micron-scale population that DLS omits.
Figure 2 (Fig. 3 in the poster). Lipid nanoparticles measured by DLS (Zetasizer Ultra, black) and Lumero (orange). The micron-scale population recovered by Lumero is confirmed in the video frames.

Viscosity is the only sample-specific input; diffusivity itself is measured even without this, and no refractive index is required. The method is validated from 20 nm to 4 µm and works 0.001-10% m/m (sample specific), using 3–30 µL per channel, collecting a data point every two minutes. Samples from transparent buffers through to turbid and strongly light-absorbing formulations have been measured as imaged. Dense or highly structured materials such as creams and pastes may still require some dilution, but dilution becomes a smaller, better controlled choice rather than a fixed requirement of the technique.

The poster presents two case studies demonstrating that dilution alters the measured sample. Lipid nanoparticles treated with triton to induce lysis, measured undiluted at 0.9 mg/mL and after ten-fold dilution in water, showed a signal reduction of approximately one thousand-fold; 100 times larger than the ten-fold reduction expected for particles robust to dilution.

Microscopy images of triton-treated lipid nanoparticles undiluted and after ten-fold dilution required for DLS, with the corresponding intensity-weighted size distributions. Visible structure in the undiluted sample is lost in the diluted one.
Figure 3 (Fig. 4 in the poster). LNPs treated with triton, undiluted (0.9 mg/mL) and 10× diluted (0.09 mg/mL), scale bar 50 µm, with the corresponding intensity-weighted distributions scaled by signal.

A liposome reference material from the National Research Council Canada, measured in its original sucrose-based buffer and after 3x dilution in the buffer specified by the certificate protocol, showed a significant increase in the proportion of large particles.

Microscopy images and normalised intensity-weighted distributions of an NRCC liposome reference sample, undiluted and after 3× dilution in PBS, revealing large particles introduced on dilution.
Figure 4 (Fig. 5 in the poster). Images and normalised intensity-weighted distributions of the NRCC liposome reference sample, with and without dilution in PBS. Here dilution increases the proportion of large particles significantly. The dotted reference value is obtained from the certification, and is specific to one model of DLS at 50× dilution.

Further examples cover verification of a filtration step, real-time growth of a DNA complex for gene therapy measured at-line, stability tests of a sunscreen formulation whose micro-scale distribution changed measurably within one hour, and quantification of large scale aggregates in a 100 mg/mL monoclonal antibody formulation after physical agitation and heat stress.

Radius of a gene therapy DNA complex plotted against time over 30 minutes, revealing a power law, beside a list of processes that can drive the change.
Figure 5 (Fig. 6 in the poster). Real-time growth of a DNA complex for gene therapy, measured at-line alongside the bulk process. The radius follows a power law in time, giving an informative growth exponent and predictive power, rather than a single end-point size. Growth rate depends on concentration, so this is only meaningful undiluted.

The Lumero S1 is an at-line and off-line benchtop instrument; within the InSPIREmed project, and with CPACT partners, Dyneval is developing an on-line implementation for direct integration into manufacturing pipelines.

Key Findings

  • A ten-fold dilution of triton-treated lipid nanoparticles reduced the signal approximately one thousand-fold. Particles robust to dilution would have given a ten-fold reduction, so the dilution changed the sample rather than merely diluting it. This is confirmed by the video evidence.
  • A certified liposome reference material gave a measurably different distribution in its original sucrose-based buffer and after a small 3x dilution in the buffer its certificate protocol specifies.
  • Lumero is applicable across a wide range of systems, many of which are demonstrated in this poster. This includes coloured, turbid, and transparent samples.
  • Measuring at working concentrations allows acquisitions to be repeated at two minutes per data point without repeated sampling for process monitoring. The radius of a DNA complex (used in gene therapy) was demonstrated to follow a power law in time, giving an informative growth exponent and predictive power rather than a single end-point size. Growth rate depends on concentration, so this is only meaningful in the undiluted sample, inaccessible to most technologies working in this size range.
Aggregate signal across four measurements of a monoclonal antibody formulation, with intensity-weighted size distributions before and after physical and heat stress showing a monomer fingerprint and aggregates on the hundreds of nanometre to micron scale.
Figure 6 (Fig. 9 in the poster). Left: aggregate signal (R > 20 nm) across the four measurements. Right: intensity-weighted distributions before and after each applied stress, showing the monomer peak and the aggregate population that grows with stress. The aggregates responsible were identified directly in the videos.

Free diffusion in water was assumed and viscosity was not quantified, so reported sizes are systematically overestimated, but signal is the important quantity. The peak below 10 nm sits outside the validated range and should be read as a reproducible fingerprint rather than an absolute monomer size, since interactions and viscosity have not been accounted for.

Authors

Picture of Dr Vincent Martinez, MPhys, PhD

Dr Vincent Martinez, MPhys, PhD

Co‑Founder & CTO, Dyneval

Picture of Dr Joe Bradley, MPhys, PhD

Dr Joe Bradley, MPhys, PhD

Lumero Product Manager, Dyneval

Picture of Dr Tiffany Wood, MPhys, PhD

Dr Tiffany Wood, MPhys, PhD

Co‑Founder & CEO, Dyneval

Cover of the Lumero poster, Particle size distributions at process or end-use conditions, presented at Mini-APACT 2026.

Want the full picture?

The poster covers ten sections in all, including filtration, concentrated monoclonal antibodies at 100 mg/mL, real-time process tracking, and early on-line flow-cell results from the InSPIREmed project.