Particle size distributions at process or end-use conditions

Poster presented at Mini-APACT 2026, AstraZeneca, Macclesfield, UK, 24–25 September 2026.

Summary

Sizing from tens of nanometres to several microns in a single acquisition using the VISTA algorithm.

A process analytic is only useful if it reports the state of the process, not the state of the sample after it has been prepared for the instrument. Most particle sizing techniques need the sample brought into a narrow optical working window, which for a concentrated process stream usually means dilution. That is not a neutral operation: self-assembled structures can be lost, osmotic changes can drive growth, and weakly bound early aggregates can be broken.

This poster sets out measurements made with Lumero S1, which records a short microscopy video of the sample in a disposable chip, using as little as 3 µL. VISTA (Video Intensity Spatio-Temporal Analysis) analyses the image intensity fluctuations arising from Brownian motion of the particles over a range of length and time scales, and returns a distribution of diffusivities, converted to hydrodynamic radius through the Stokes–Einstein relation. Because the analysis is statistical rather than single-particle, particles need not be resolved or tracked, and billions can be captured in one measurement.

Single frame from a Lumero acquisition of a lentiviral vector sample, shown beside its intensity-weighted and volume-weighted size distributions which reveal coexisting small and large particles.
Figure 1 (Fig. 1 in the poster). A single frame from a Lumero acquisition. Only the largest particles are individually visible; the sub-resolution population still modulates the image intensity and is measured by VISTA even though it is invisible to manual inspection, and the spatial localisation of large particle signal means the small signal can still be detected.
Intensity-weighted particle size distribution of lipid nanoparticles measured by DLS and by Lumero, showing a micron-scale population that DLS omits.
Figure 2 (Fig. 3 in the poster). Lipid nanoparticles measured by DLS (Zetasizer Ultra, black) and Lumero (orange). The micron-scale population recovered by Lumero is confirmed in the video frames.