Particle size distributions at process or end-use conditions
- Dyneval
Poster presented at Mini-APACT 2026, AstraZeneca, Macclesfield, UK, 24–25 September 2026.
Summary
Sizing from tens of nanometres to several microns in a single acquisition using the VISTA algorithm.
A process analytic is only useful if it reports the state of the process, not the state of the sample after it has been prepared for the instrument. Most particle sizing techniques need the sample brought into a narrow optical working window, which for a concentrated process stream usually means dilution. That is not a neutral operation: self-assembled structures can be lost, osmotic changes can drive growth, and weakly bound early aggregates can be broken.
This poster sets out measurements made with Lumero S1, which records a short microscopy video of the sample in a disposable chip, using as little as 3 µL. VISTA (Video Intensity Spatio-Temporal Analysis) analyses the image intensity fluctuations arising from Brownian motion of the particles over a range of length and time scales, and returns a distribution of diffusivities, converted to hydrodynamic radius through the Stokes–Einstein relation. Because the analysis is statistical rather than single-particle, particles need not be resolved or tracked, and billions can be captured in one measurement.